Page 81
6.
Put a few drops of oil on the two door pins and door tightening bolts.
1.
Clean outer parts of the autoclave with a soft cloth.
11.1.3
Periodically
1.
Every 6 months replace the air filter, (if installed) (to be done by an
authorized technician).
2.
Once a year check and tighten the piping joints to avoid leakage (to be
done by an authorized technician).
3.
Once a year check and tighten all screw connections in the control box,
heaters, valves and instrumentation (to be done by an authorized
technician).
4.
Check the door gasket every 12 months and replace it if required (refer
to the technician manual). Replacing the gasket shall be done by a
technician.
11.1.4
Periodical Tests
1.
Once every month activate the safety valve (see sec. 12.2).
11.2
Checking the Safety Valve
The safety valve is located on the rear side of the autoclave
In order to prevent the safety valve from a blockage, operate it once a
month.
11.2.1
PED-approved type safety valve
Operate the sterilization cycle according to the manual.
Allow a pressure of approximately 200 kPa (29 psi) to build up in the
chamber.
Turn the safety valve pressure-regulating nut clockwise for 2 seconds.
Be careful not to burn your hands.
Return the nut to its original position.
Press the
STOP/STOP
key to interrupt the operation, and exhaust
steam from chamber.
Wait until the pressure decreases to zero, only then can the door be
opened.
Attention:
Use protective gloves in order not to burn your hands with the hot
steam.
Содержание 2840 ELV-D
Страница 21: ...Page 20 3 15 Overall Dimensions Drawing mm...
Страница 74: ...Page 73...
Страница 80: ...Page 79 To avoid injuries use heat resistant gloves while unloading the autoclave...
Страница 84: ...Page 83 Attention Use protective gloves in order not to burn your hands with the hot steam Safety valve ring...