Page 44
5.12
Program 12: Isothermal
For liquids when the manufacturer recommends autoclaving at 60-
100
C (140-212
F) with no drying. Recommending for melting of agar,
pasteurization etc.
Caution!
This is not a sterilization program!
Put one PT100 inside the bottle, leave the second one hanging in the
chamber outside the bottle (see below).
Nominal parameters default settings
•
Heating temperature (default): 80
C (176
F).
•
Heating time: 20 minutes.
Operations sequence:
•
Water Inlet:
Water enters the chamber