Page 14
Shipping volume
0.56 m3
Shipping weight
115 kg
Shipping
dimensions
length
730mm
width
610mm
height
125mm
Max. Allowable Working
pressure (MAWP)
2.8 bar
Chamber
diameter
280mm
Depth
400mm
Chamber Volume
31L
3.9
Construction
Chamber and door material
Stainless steel
Outer Cabinet
Stainless Steel
Chamber insulation
Fiberglass with reinforced material
3.10
Utilities
Power Supply
Power
1 Ph, 230V/50/60Hz
Recommended
Circuit Breaker
16A
Compressed Air (ELVC only)
1/2" 3 Bar (44-58 psi)
Tap water
1/2", 2-6 Bar (29-44 psi)
Mineral free water
1/2", 2-3 Bar (29-44 psi)
Drain
2" Minimum Withstanding temp. of 80 M
(17676)
Attention!
•
A switch or circuit-breaker must be included in the building
installation. This switch or circuit-breaker shall be in close
proximity to the equipment, within easy reach of the operator; and
marked as the disconnecting device for the equipment.
•
The electrical net must be protected with a current leakage safety
relay.