Page 37
Problem/
Error
Message
Message / Symbol Description
Corrective Action
"Door is open
(During the
cycle)"
This message is displayed when the
door is open: During the cycle.
Close the door to perform a
new cycle.
"Canceled
By User"
This message is displayed after the
START/STOP key is pressed and
cycle aborted.
Wait until "cycle failed –
canceled by user" or "cycle
end – canceled by user" is
displayed. Perform a new
cycle.
"Air Error"
This message is displayed at the end
of the cycle If the autoclave does not
reach the atmospheric pressure after
10 minutes.
Wait until the autoclave
reaches the atmospheric
pressure and perform a new
cycle.
"Mineral free
water
reservoir
empty"
This message is displayed if the
water level switch in the reservoir
does not sense water.
Fill the mineral free water
reservoir (see 6.4).
"Power
Down"
This message is displayed if power
down has occurred during the cycle.
(this message will print out in the
printer after the autoclave will turn
on).
Turn on the autoclave and
wait until the autoclave is
ready (reaches the safe
condition) and perform a new
cycle.
"No Water"
This message is displayed if the
electrode in the chamber did not
sense water within the preset time.
Check and fix the mineral free
water
supply.
Check and clean the water
inlet filter.
Clean the water level
electrode.
The following messages require you to check that your autoclave is not
overloaded:
"Heat Time Error"
"Heat Time Error (Keep)"
"Pressure Time Error"
The following messages require you to perform a new cycle:
"Low Pressure"
"High Pressure" "
"High Pressure (Ending)"
"High Pressure (Exhaust)"
"High Pressure (Dry)"
"Cycle Failed"
\