Page 34
10.2. Draining the Reservoir
Caution!
Before starting, Make sure that the electric cord is
disconnected and there is no pressure in the autoclave.
It's the user's duty to drain the reservoir once a week. The drain valve is
located on the front right side of the autoclave after the door is opened. The
function of the drain valve is to drain the water reservoir.
1.
To drain the reservoir, use item (5) with the plastic hose (6) attached to
it (supplied with the autoclave).
2.
Insert part (5) into valve (3) and press it until you hear a click. The drain
valve opens immediately, drain into a bucket.
3.
When the water reservoir is empty, press part (4). Item (5) will pop out
approx. 3mm and the drain valve will be closed. Remove item (5) with
the plastic tube.
4.
Fill reservoir with distilled water until it reaches the full level (see 6.4).
5.
Turn on the main power switch.
6.
The autoclave is now ready for use.
2340EA-D, 2840EA-D
3
3850EA-D, 3870EA-D