Navigation through the menus is done simply by touching the item of interest. After performing an
adjustment you will return to the main home screen by touching menu. You will find that lightly
tapping with your fingernail works the best. You will hear a beep when a selection is made.
Touching the Menu button will display the main submenus available to configure the
analyser
.
The software is designed to provide simple operation, therefore some functions and or
capabilities are programmed at the factory and are not field adjustable.
Displays and sets the operating range of the
analyser
. Setting the range
automatically scales the 4-20 mA output to the selected range and is not further scalable.
Setting the range also automatically sets the high and low alarm set points to 10% and 90% of
the range. The alarm trip points can be configured to other values.
Is used to set the high and low disinfectant alarm levels; and to set a
time delay for each alarm before it will energize. Alarms are deactivated during calibration.
Is used to perform the zero point calibration if needed and span
calibration to a grab sample or standard.
These are used to adjust screen brightness and audio volume and to
set pH compensation to manual or auto if this feature is available (for systems with pH sensor
only). This screen also contains the serial number, the software version number, and display
version number.