Screw the terminal cover onto the sensor and then tighten the cable gland
nut.
The solution ground pin on the front of the flow cell is normally not needed,
it may be used if you find a solution ground is necessary.
The sensor will not provide a constant value until after a conditioning time of 2 hours has
passed. It cannot be calibrated until then.
The calibration procedure should be repeated on the day after initial commissioning.
On the chlorine sensor, the screw coated with thread locking paint must not be
adjusted. If the thread locking paint is damaged claims under the warranty will not be
honored.