Free Chlorine Sensor Data, 2 -Electrode & 3-Electrode with Reduced pH Dependence
Analyte
Membrane type
Hydrophobic PTFE membrane
Hydrophilic membrane
Measuring cable
connection
2-pin terminal, polyamide PG7 screw connection;
wire cross section 2x 0.25mm2, cable diameter approx. 4 mm
Suitable chlorination agents
Inorganic chlorine compounds: NaOCl (sodium hypochlorite), Ca(OCl)2,
chlorine gas, chlorine produced by membrane electrolysis (not suitable:
chlorine electrolysis without a membrane)
Voltage supply
UB 12 to 30 V DC (electrical isolation recommended)
Electromagnetic
compatibility
According to EN 61326-1
Interference emission: Class B
Interference immunity: To industrial requirements
Output signal
4 to 20 mA
Settling time
1 h o u r
2 h o u r s
Incident flow velocity
Approx. 15cm/s or 30L/hr (8 GPH)
Measuring ranges mg / l (ppm)
0.05 to 0.5
0.05 to 2
0.05 to 5
0.05 to10
0.05 to 20
0.05 to 50
0.05-100, 0.05 to 200
0.05 to 2
0.05 to 5
0.05 to10
0.05 to 20
0.05 to 200
Resolution
0.01 mg/l for 0.5,2,5,10,20 mg/l
0.1 mg/l for 50, 100, 200 mg/l
0.01 mg/l for 2,5,10,20 mg/l
0.1 mg/l for 50, 100, 200 mg/l
Response time t90
About 30 sec.
About 2 min
Operating temperatures /
temperature compensation
+5 to +45 °C
Zero point adjustment
Not required
pH value operating range
6.0 to 8 pH
4 to 9 pH
pH dependency
(loss of slope)
No loss from pH 5-7, at pH 8 about
65%, at pH9 about 90%
No loss from pH 5-7, at pH 8 about
10%, at pH9 about 20%
Disruptive substances /
cross sensitivity
Chlorine dioxide, Ozone, Combined
Chlorine, reducing agents,
corrosion protection agents,
Calcium deposits.
Chlorine dioxide, Ozone, Combined
Chlorine. Calcium, Iron over
USEPA MCL prevents electrode
function. Corrosion inhibitors
Min. Sample Conductivity
50µS / cm (not suitable for highly purified water)
Pressure resistance
pabs max. 2 bar
prel max. 1 bar
No pressure fluctuations are admissible when operating under pressure.
We recommend unpressurized operation (atmospheric pressure).
Material
Shaft, cover, cap: PVC
Shaft, cover, cap: PVC, Membrane
disk holder: stainless steel
Dimensions & Weight
Diameter: 25mm, length: 220mm, Weight about 125 g
Maintenance
Check the measurement signal: regularly, at least once a week
Replace the membrane cap: once a year (subject to water quality)
Change the electrolyte:
every 3 to 6 months
Storage
Sensor: frost-free, dry and without electrolyte, can be stored for an
unlimited time at +5 to +45 °C
Electrolyte: In the original bottle and protected against sunlight at +5 to
+25 °C