The service life of the electrolyte is 3 to 6 months. The service life of the membrane depends
on the water quality.
If the sensor still indicates values that are too low after the electrode finger tip has been cleaned, a
new membrane cap must be used.
Unscrew the membrane cap from the shaft of the sensor. See "Screwing the
membrane cap off and on".
Flush away the electrolyte.
Place the membrane cap in a 1% solution of hydrochloric acid for a few hours.
Before commissioning rinse off with distilled water.
Screw the membrane cap filled with electrolyte onto the shaft of the sensor.
Unscrew the membrane cap from the shaft of the sensor see "Screwing the
membrane cap off and on".
Flush away the electrolyte.
Rinse off the membrane cap and electrode finger with distilled water and dry
with a dust free cloth.
Screw the dry membrane cap loosely onto the electrode shaft. The
membrane must not be resting against the electrode tip.
Used membrane caps that have been in operation for longer than 1 day
cannot be reused!
Clean the electrode tip with the enclosed special abrasive paper, see
"Cleaning the electrode finger tip / replacing the membrane cap and
electrolyte".
Fill the membrane cap with electrolyte that is free of air bubbles.
Use a new membrane cap.