Inspect the equipment and packing materials for signs of shipping damage. If there is any evidence of
damage, notify the transit carrier immediately.
After unpacking, it is recommended to save the shipping carton and packing materials in case the
instrument must be stored or re-shipped.
The shipping container holds the following:
1-
TT-ISE-CL2
Amperometric Chlorine analyser
1- Chlorine sensor with 100ml bottle of electrolyte and abrasive cloth
1-Measuring cell
1 -Sample Tubing 8mm OD x 6mm ID x 5’ long, clear flexible PVC tubing
1- Sample Drain Tubing 8mm OD x 6mm ID x 5’ long, clear flexible PVC tubing
1- Grab Sample Tubing 8mm OD x 6mm ID x 3’ long, clear flexible PVC tubing
1- Set of Four Mounting Feet with Four Screws
1- Power cable, 2-meter long with IEC 60320 C13 & NEMA 5-15P connectors
1- Instruction Manual
Panel mounted analyzers will include a 1/2” x 12” x 24” PVC panel to which the following is mounted,
wired and plumbed: chlorine analyser, flow cell, sample input ball valve, pressure regulator with
gauge, 12 GPH flow meter, 3-way grab sample valve, sample and drain tubing. No mounting feet
included.
The
TT-ISE-CL2
chlorine analyser is shipped with two serial number labels found on the bottom right
side of the enclosure, and on the inside of the enclosure door. The labels include the factory
calibrated chlorine residual range. Should technical assistance be required, refer to the serial number
to identify your system.
The serial number and software version number can also be found on the "Settings" main screen.
This number is required if technical support service is needed.
Analyser
The
TT-ISE-CL2
Amperometric Reagentless Chlorine Analyser is designed to be mounted indoors on
a vertical wall or panel. Mount in a location that is:
• Within ambient temperature and humidity limits (41-113°F, +5 to +45°C), 95% relative
humidity or less, non-condensing
• Out of direct sunlight and away from any heat source
• Protected from direct contact with water, rain, snow, ice, or dust.
• Protected from corrosive fluids
• An area that is not subject to vibration
• As close to the sample point as possible to minimize sample lag time
• Accessible for viewing and maintenance
If the analyser must be located outdoors it must be mounted in a separate enclosure that provides
protection from direct sunlight and excessive temperatures, direct contact with water, rain, snow,
ice, or dust.