1.
Make sure the sample inlet ball valve is closed, the flow meter valve is closed and the drain
valve is open. The pressure regulator is set to 15 PSI at the factory.
2.
Turn on the
analyser
by pressing the off/on rocker switch located on the bottom right
underside surface of the
analyser
enclosure.
3.
The gray splash screen will appear as the software boots up,
followed by the Home
operating screen.
On the circuit board the heartbeat indicator below the processor will be flashing red, and the “Power
On” red LED indicator will light.
The screen will display the current sample residual and water temperature. If the pH measurement
option is included the sample pH will also be displayed. The heartbeat indicator will flash in the
lower left corner.
4.
Slowly open the sample inlet ball valve to deliver sample to the system.
5.
Make sure the water pressure gauge reading on the pressure reducing regulator is no more
than atmospheric pressure, 15 psig. To adjust the pressure regulator:
a. Loosen the check nut. Turn the adjusting screw clockwise to raise the reduced
pressure, turn counter clockwise to lower the reduced pressure. Do not bottom out the
adjusting screw.
6.
With the water pressure reduced, slowly open the flow meter valve by turning the knob
clockwise until a flow rate is at least 8 GPH. Flow may safely be set up to 12 GPH without
affecting the chlorine sensor readings.
7.
Make sure there are no pressure or flow fluctuations.