![Turtle Tough TT-ISE-CL2 Скачать руководство пользователя страница 23](http://html1.mh-extra.com/html/turtle-tough/tt-ise-cl2/tt-ise-cl2_manual_1182459023.webp)
TURTLE
TOUGH
®
23
TT-ISE-CL2
General
Product Description
Amperometric reagentless total chlorine residual analyser
Intended Use
Continuous monitoring and control of chlorine residual prior to dechlorination
Measurement Method
Amperometric potentiostatic membrane covered 3-electrode
Parameters Measured
Total (combined) residual chlorine
Available Operating Ranges
0-0.5 mg/l to 0-20 mg/l (ppm)
Temperature Compensation
Standard, +5 to +45
°C
Mechanical
Dimensions
15.25” W x 12.468” H x 6” D, bottle bracket 6.75” dia.
Weight Approx.
15 lbs.
Enclosure & Ingress Protection
Wall mount molded fiberglass, UV resistant, NEMA 4X, IP65 Equivalent
Ambient Temperature and Humidity
+5 to +45
°C
; 0-95% relative humidity, non-condensing
Electrical
Conduit Openings
2
Operating Power & Consumption
24 VDC; less than 3 watts
Power Supply
Switching 88-264 Volts AC, 50/60 Hz. Input. Single phase. 24 VDC 2.2A, 52.8W
Power Input Connection
External 6A fused module, IEC 320-C14 connector, with SPST switch, 2 meter
detachable cord with IEC 60320 C13 & NEMA 5-15P connectors
Input Fuse
6 amp, 125V, fast acting, 5 x 20 mm glass tube, Designed to UL/CSA 248-14
Electronics
Microprocessor based
Touch Screen Display / Interface
4.3” resistive full color LCD, LED backlight, 65,356 colors, screen resolution 480
x 272, 4MB flash memory, rated min. 1 million touches on any one point
Inputs
(4) Digital, (4) 4-20mA
Signal Output (non isolated)
4-20mA DC 750 ohm max load, (1) residual, up to (4) optionally
Serial Output (non isolated)
(1) RS485 serial port, non-isolated, inactive
Relays
1-amp single pole form C, (3) standard for high & low chlorine; no flow alarms.
Up to (8) available optionally. Each relay with red LED energized indicator
Sample Requirements
Sample Supply
Continuous consistent flow, electrodes must be kept wet, under power and
exposed to chlorine
Flow Rate
30L/hr (8 GPH)
Sample Pressure
5 psig, 15 psig maximum
Sample Temperature
+5 to +45 °C
Sample pH Range
4 to 12 pH, Linear slope decrease with approx. 5 % per upward pH unit
(starting from pH 7)
Mechanical
1/2” x 12” x 24”
9” Wide x 12” High x 6” Deep
Mounting Panel Dimensions
analyser Enclosure Weight with
Mounting Panel
19 lbs
Содержание TT-ISE-CL2
Страница 2: ...TURTLE TOUGH 2...
Страница 34: ...TURTLE TOUGH 34 8 9 10 analyser...