TD1000C™Oil in Water Monitor
CALIBRATION
PN 101687
REV G
TD-1000C Oil In Water Monitor
Operation Manual
21
ENT
CALIBRATION SO-
LUTION
9. Inject 120 ml of calibration solution (flush with 60 ml, then allow portion of
second 60ml to remain in flow cell). Press <
ENT
>.
ENT
CAL SOLN: XXX
10. If the number shown is between 1-10, wait for reading to stabilize, then
press <
Ý
>. WAIT command will appear.
Ý
CAL SOLN: XXX
wait/WAIT
11. When finished, monitor will prompt you to press <
ENT
> to continue.
To Continue Press
<ENT>
12. Press <
ENT
>.
ENT
Press <1> to End Calibration
13. Press <
1
> to accept the new calibration settings. Monitor returns to
calibration menu. Open valve to start sample flow by returning valve handle to
the vertical position. Calibration is finished.
5.6 CELL CALIBRATION SCREENS
HOME
XXXX RFU T:
23C Cell: 100%
1. From the HOME Screen press <
¾
>
¾
CELL CALIBRATION <ENT> TO
START
2. Press <
ENT
> to initiate the cell calibration sequence
ENT
FLOW CELL CLEAN?
<ENT> TO CONT.
3. Press <
ENT
> if the flow cell is clean. Press <
HOME
> and clean flow cell if
flow cell if necessary. Restart cell calibration after cleaning flow cell.
ENT
PROCESS WATER
4. Press <
ENT
> Open inlet process water valve (handle positioned vertically).
Verify outlet valve is open to process (handle is vertical). If your instrument has
PVC wetted parts, position the outlet valve handle downward.
Ý
CELL CALIBRATION
X.XXXV
Press <
Ý
> when the measured process water measurement is stable.
1
PRESS <1> TO END CALIBRA-
TION
5. Press <
1
> to accept the cell calibration.