TD1000C™Oil in Water Monitor
PN 101687
REV G
TD-1000C Oil In Water Monitor
Operation Manual
23
positioned when the green valve handle is orientated horizontally.
If your instrument has PVC wetted parts, position the inlet valve handle horizontally and the
outlet valve in the drain position, with the end of the valve handle pointed up. See Appendix K.
3. Remove the clean-out plug.
After injecting cleaning solution into the flow cell, be sure to flush it out completely BEFORE removing
the clean-out plug.
4.
Dip flow cell brush into cleaning solution and insert gently into clean-out opening.
5. Slide brush gently up and down in the opening to remove any coating on the quartz cuvette.
6.
Replace flow cell clean-out plug.
The flow cell is not guaranteed against cracking caused by overtightening the clean-out plug. If your
flow cell features a bulk-head type fitting, use two wrenches to avoid over torquing.
7.
Open the flow-cell outlet valve to the process flow position. The valve is correctly positioned
when the green valve handle is orientated vertically. Open the flow cell inlet valve. The valve is
open when the yellow valve handle is oriented vertically.
If your instrument has PVC wetted parts, position the outlet valve handle downward and the
inlet valve handle vertically. See Appendix K.
8. Turn on the Main Power Switch.
9. Calibrate the monitor after allowing it to warm up for at least 15 minutes.
6.3 REPLACEMENT PARTS
See
Appendix H
for information on common replacement parts.
Improper installation, use, application or unauthorized servicing of this equipment voids all
guarantees.
7 RESPONSIBILITY FOR SAFE DELIVERY
Turner Designs Hydrocarbon Instruments has done everything possible to protect this equipment
from damage due to normal transportation hazards. After the product leaves the manufacturing site,
responsibility for its safe handling and delivery is assumed by the transportation company handling the
equipment.
If the crated monitor shows evidence of rough handling, request that the person making the delivery
write “Received In Damaged Condition” on the delivery receipt. If concealed damage is revealed after
the shipment is unpacked, contact a representative of the Transportation Company and request that a
“Damaged Goods” report be completed.
In either event, the Transportation Company should be notified immediately of any damage to the
shipment to protect your rights of recovery.