START-UP
TD1000C™Oil in Water Monitor
10
TD-1000C Oil In Water Monitor
Operation Manual
PN 101687
REV G
4.1 START-UP PROCEDURE
A diagram of the screens is located in Appendix F.
STEP 1
Bring sample flow to and from monitor - check for leaks in plumbing.
STEP 2
Switch Main Power Switch under enclosure face to the ON position (see
Figure 1
).
When power is first turned on, an ID screen will appear for a few seconds, showing the software version
and date. After 10 seconds (or press <ENT> or <HOME> for immediate access), the HOME screen will
appear. The HOME screen will display the RFU value (instruments with software 103271 REV B and
newer will display PPM as the default unit), the process water temperature and the Cell Condition %.
x.x RFU
T: xxC CELL: xxx%
Sample Home Screen
Monitors with newer firmware will read PPM instead of RFU.
If the screen is blank or dark, try adjusting the screen contrast using the UP <
ñ
> and DOWN <
ò
>
arrows.
STEP 3
During start-up, eight system values must be entered. Record all set-up values in the configuration
record located in
Appendix B
.
Press <ENT>. Before entering/changing the first System Value, the monitor will prompt for the User ID.
The screen will read:
PLEASE INPUT ID:
Enter a valid USER ID (default is 0258) and the screen will display:
CURRENT: 0258
NEW: