TD1000C™Oil in Water Monitor
START-UP
PN 101687
REV G
TD-1000C Oil In Water Monitor
Operation Manual
13
<
2
>
m
inuteS
•
Press <
¹
>, from HOME.
•
Press <
2
>.
•
Key in minutes [0–59].
•
Press <
ENT
>.
•
Press <
ESC
> to return to clock menu.
<
3
>
m
ontH
•
Press <
¹
>, from HOME.
•
Press <
3
>.
•
Key in the month (1–12).
•
Press <
ENT
>.
•
Press <
ESC
> to return to clock menu.
<
4
>
D
ay
•
Press <
¹
>, from HOME.
•
Press <
4
>.
•
Key in the day (1–31).
•
Press <
ENT
>.
•
Press <
ESC
> to return to clock menu.
<
5
>
y
ear
•
Press <
¹
>, from HOME.
•
Press <
5
>.
•
Key in year [00-99].
•
Press <
ENT
>.
•
Press <
ESC
> to return to clock menu.
After start-up, wait a minimum of 15 minutes before Cell Condition Calibration to allow the monitor to
come to equilibrium.
STEP 5
Perform a site specific Cell Condition calibration. See
Section 5.1
for the steps involved in performing
this procedure.
The TD1000C™ monitor start-up has now been completed.