TTI PSA1301T Скачать руководство пользователя страница 35

  

  34 

20.2.2.  Centre Menu Shortcuts 

Set Centre to Marker:

 To re-centre the sweep to a new position within the currently visible sweep, move 

the marker (M1) to the desired position and press Set C = M1. 

Step Up / Step Down:

 The centre frequency can be moved in fixed steps (settable between 1kHz and 

100MHz) by these two keys.  The Set Step Val. key accesses the menu for setting the step size. 

Step Value Shortcuts:

 

The Step Value sub menu provides two shortcut keys.  Step = MΔ sets the step 

size to the difference between the two marker positions M1 and M2.   Step = Span/10 sets the step size 
equal to one tenth of the current span i.e. one graticule division. 

20.2.3.  Span Menu Shortcuts 

Zoom In / Zoom Out:

  The span can be increased or decreased in binary steps by these keys. 

Set Span equal to Marker Difference:

 

The Spn = MΔ key sets the span equal to the difference between 

the two marker positions M1 and M2 while moving the centre frequency to the markers mid-point. 

20.2.4.  Start and Stop Menu Shortcuts 

Set Start and/or Stop to Markers:

 The Start = M1 key moves the start frequency to the current position 

of marker M1.  An equivalent functions is available from the Stop menu.  The Strt=M1/Stp=M2 key sets 
the start and stop frequencies to the current marker positions. 

20.2.5.  Setting by Direct Numeric Entry 

Each frequency related menu has a Set by Keybd key.  This creates a 0 to 9 numeric entry keypad of 
large keys which can be pressed directly with the finger tip.  (Alternatively they can operated with the five 
way navigator).  The entered value appears at the top. 

The space below the keys contains the standard Palm input area.  If required, this can be used to edit 
the number after it is entered and before it is confirmed using standard Palm cursor-based editing. 
See section 20.1 for details of how to change the input area type. 

Press OK to activate the new frequency and return to the menu, or Cancel to abort and return to the 
menu.  Note that the Back (Home) hard key does not operate within this screen. 

If Set by Keybd was the last function used within a menu, then it will be automatically selected the next 
time that the menu is entered.  This reduces the number of key presses required when entering new 
frequencies repetitively. 

20.2.6.  Setting by Tab and Jog 

Setting by Tab/Jog provides an alternative to settings frequencies by the Keyboard.  It is particularly 
relevant to users who wish to operate the PSA1301/2701T without using the touch screen, but it is 
potentially relevant to all users. 

Within any frequency menu, the current set frequency is displayed above the keys in a set of boxes.  
The Set by Tab/Jog function allows the frequency to be changed .using the 5-way navigator (left/right to 
select digit, up/down to change digit).  It is also possible to select the digit to be changed simply by using 
the stylus or fingertip to press it. 

This also provides a quick and convenient way of stepping frequencies (although the step size is limited 
to decade values). 

New frequencies do not have to be confirmed.  A delay of about half a second is included between a 
digit being altered and the sweep parameters being changed to match.  This allows digits to be scrolled 
without the sweep continually restarting. 

If Set by Tab/Jog was the last function used within a menu, then it will be automatically selected the next 
time that the menu is entered.  This reduces the number of key presses required when entering new 
frequencies repetitively. 

Содержание PSA1301T

Страница 1: ...THURLBY THANDAR INSTRUMENTS INSTRUCTION MANUAL PSA T Series PSA1301T PSA2701T 1 3GHz 2 7GHz RF Spectrum Analyzers...

Страница 2: ...ency Measurement 8 6 2 Amplitude Measurement 8 6 3 Sweep 9 6 4 Signal Input 10 6 5 Demodulation Zero Span mode 10 6 6 Display 10 6 7 Memory Storage 10 6 8 Connectors 11 6 9 Power Sources 11 6 10 Mecha...

Страница 3: ...2 18 2 Selecting and Leaving the Application 22 18 3 Touch Screen or Hard Key operation 22 18 4 The PSAnalyzer Screen 23 18 5 Operation and Navigation 23 18 6 Turning Off Power Saving 24 19 The PSAnal...

Страница 4: ...sing a printer connected to a PC 40 22 3 Transferring PSAnalyzer Files to a Personal Computer 40 23 PSAnalyzer Files and Defaults 41 23 1 Factory Default Settings for PSAnalyzer 41 23 2 File Locations...

Страница 5: ...ch are shown as follows see section X X Within a PDF file the boxed number is a hyperlink to that section number thus enabling the user to jump rapidly the section referred to and then jump back to co...

Страница 6: ...primary function of RF spectrum analysis without linkage to other devices TTi provides a USB linked memory card reader in order that PSAnalyzer files can be transferred to a PC where this is required...

Страница 7: ...d key Some general information regarding the capabilities of the unit is available from the Info key on the Exit menu 4 2 Detailed Operating Guide To fully understand the operation of the PSA1301 2701...

Страница 8: ...cally peak finding and tracking A reference trace can be displayed in addition to the live trace Limit lines can also be created and displayed The graticule is fully controllable Any number of traces...

Страница 9: ...ppm over 10o C to 30o C Ageing Better than 3 ppm per year Phase Noise Phase Noise Typically 100dBc Hz at 100kHz offset at 500MHz PSA1301 Typically 90dBc Hz at 100kHz offset at 500MHz PSA2701 Resoluti...

Страница 10: ...age 55dBc typically 60dBc Residual Spurii Below noise floor PSA1301 3dB above noise floor PSA2701 Amplitude Markers Number of Markers One Two or None Marker Resolution 0 1 dB Amplitude Readout Display...

Страница 11: ...Stored trace either recalled from memory or copied directly from live trace Selectable on or off Resolution Modes Selectable as High Resolution or Low Resolution In low resolution mode the trace area...

Страница 12: ...from mains power using the AC power adaptor provided by TTi This powers and recharges both the Spectrum Analyzer and the Palm T X simultaneously AC Adaptor Charger Input Voltage Range 100V to 240V no...

Страница 13: ...use which angles the unit at approximately 25 degrees to the horizontal and can alternatively be used as a hook mount 6 11 Environmental and Safety Operating Range 5 o C to 40 o C 20 to 80 RH Storage...

Страница 14: ...C Compliance was demonstrated by conformance to the following specifications which have been listed in the Official Journal of the European Communities EMC Emissions a EN61326 2006 Radiated Class B Im...

Страница 15: ...e susceptibility to high electromagnetic fields throughout their measurement range In general the instrument will operate correctly but may show the interfering frequencies on the measurement trace if...

Страница 16: ...that the shell of the mini USB connector is galvanically connected to the body of the SMA input which will therefore be at earth ground potential when the USB port is connected to a desktop PC To main...

Страница 17: ...echarging When they are fully charged this LED will go off The charging state and charge level of the Palm T X batteries is shown on the display of the T X see section 17 1 10 3 Demodulated Audio Outp...

Страница 18: ...available The memory card can be used as a means of transferring data to and from a PC using the USB card reader supplied See section 22 3 for details 12 1 Removing and Inserting the Card The card fi...

Страница 19: ...Removing Re fitting Remove the Stylus from the Palm T X Push the Palm T X onto the connector by pushing it firmly in the line of the instrument Ensure that the two fixing latches are open Locate the t...

Страница 20: ...tialise 14 2 Technical Note about the OFF state Handheld computers operate on the principle of being always on When they are turned off they merely go into a low power consumption mode with the displa...

Страница 21: ...Pressing it again turns it back on Note It may be necessary to press the key more than once to turn on because of slow key polling in sleep mode The PSAnalyzer application suppresses automatic switch...

Страница 22: ...hours of continuous operation from a charge Recharge time is typically less than four hours To conserve the battery the PSAnalyzer application can be put into sleep mode when not in use by pressing a...

Страница 23: ...r and turn the function off The USB lead should be unplugged at the spectrum analyzer end The USB cable could induce interference into the spectrum analyzer that could cause spurious signals to be dis...

Страница 24: ...Screen Data Area 2 Icon Area Data Area 1 Primary Frequency Settings Reference Level Resolution B W Video Filter Sweep Run Stop Battery Status Data Area 3 Data Area 4 Marker Readout Sweep Type Attenuat...

Страница 25: ...o Span or Quick Menu Also provides a function called Preset which sets defined conditions from a single key press Note that the labels of the next two keys depend upon the mode setting Centre set the...

Страница 26: ...jacent to the RF Signal Input Demodulation mode and audio level are selected from the AM FM menu Quick Menu is a different mode of operation which provides direct access to some of the most commonly u...

Страница 27: ...and return to the home menu press the Back Home hard key or the grey area at the top left of the menu 19 5 Step Value Sub Menu There are four ways to set the step value Set Step M delta sets the step...

Страница 28: ...y area at the top left of the menu 19 8 Stop Frequency Menu There are three ways to set the stop frequency Stop M1 sets the stop frequency equal to the frequency at the position of Marker One Start M1...

Страница 29: ...B range To exit this menu and return to the home menu press the Back Home hard key or the grey area at the top left of the menu 19 11 Markers Menu Turn Markers On press the centre of the 5 way navigat...

Страница 30: ...hs of 280 kHz and 15 kHz PSA1301 or between the three resolution bandwidths of 1 MHz 280 kHz and 15 kHz PSA2701 Normal sets a continuous sweep mode where a new sweep is commenced automatically after t...

Страница 31: ...limit lines have been created or recalled the key will have no function The colour of Limit Line 1 is red Limit line 2 is blue To exit this menu and return to the home menu press the Back Home hard k...

Страница 32: ...ered to create limit lines for the display and to save and recall limit lines from memory see section 20 9 To exit this menu and return to the home menu press the Back Home key or the grey area at the...

Страница 33: ...Marker control and no longer operate the key highlighting The purpose of Quick Menu is to provide direct access to some of the more commonly used functions without having to select sub menus Frequenc...

Страница 34: ...enus see section 19 Moving between menus requires a return to the Home menu by pressing the Back Home hard key When using the touch screen for control rather than the 5 way navigator it is also possib...

Страница 35: ...equired this can be used to edit the number after it is entered and before it is confirmed using standard Palm cursor based editing See section 20 1 for details of how to change the input area type Pr...

Страница 36: ...sweep and the old sweep values are sequentially over written with the new values In this mode the progress of the sweep is indicated by a moving yellow line at the bottom of the graticule Video Filter...

Страница 37: ...s described in section 19 11 Markers are vertical lines with a box that highlights the intersection point of the marker with the trace The readout shows the frequency and level at the intersection poi...

Страница 38: ...o store traces and complete instrument set ups to a memory card and recall them as required See Stores menu section 19 14 Each key within the Stores menu brings up a dialogue box which includes a scro...

Страница 39: ...is deletes all points in the current list To exit the editor screen while saving all of the changes press the Finish key To exit without saving any changes press the Cancel key Note that frequency amp...

Страница 40: ...r measurement accuracy will be achieved using Hi Res Mode 20 11 Changing the Palm Input Area type The Set by Keyboard function for frequency setting and all Store Recall functions brings up the standa...

Страница 41: ...anywhere on the screen will return to the thumbnails view Press the home key to exit the Media application 22 2 Printing a Stored Screen Image using a printer connected to a PC Screen image files are...

Страница 42: ...pen the disk drive under its allocated drive either from My Computer or from Explorer The directory structure of the card will be visible Trace files are stored in the folder PSAnalyzer Tr21aces Set u...

Страница 43: ...s 120 kilobytes and is an image of 379 pixels high by 320 pixels wide in 8 bit indexed colour format When transferred to a personal computer screen files can be opened in any image editing program and...

Страница 44: ...4 4 Calibration To ensure that the accuracy of the instrument remains within specification the calibration should be checked and if necessary adjusted annually The calibration adjustment procedure is...

Страница 45: ...er button 2 While holding the power button use the tip of the stylus or a similar object without a sharp tip to gently press and release the reset button 3 Wait for the Palm Powered logo to appear not...

Страница 46: ...Shortcut to PSAnalyzer As supplied by TTi the PSAnalyzer application is installed within the Palm T X applications so as to occupy the first position within the Applications screen category All or Unf...

Страница 47: ...idgeshire PE29 7DR England United Kingdom Telephone 44 0 1480 412451 Fax 44 0 1480 450409 International web site www tti test com UK web site www tti co uk Email info tti test com Aim Instruments and...

Отзывы: