TTI PSA1301T Скачать руководство пользователя страница 20

  

   

 

19 

14.

 

 Re-initialising the Palm T|X 

Whenever the battery of the Palm T

|

X has been allowed to completely discharge (as is normally the 

case when it is first received) it will be necessary to re-initialise it. 

14.1.  Procedure for Re-Initialising 

Charge the Palm T

|

X in the normal way.  After a short period (normally less then 10 minutes) the battery 

voltage will recover to a point where normal operation is possible.  The T

|

X may then switch itself on 

automatically, or it may be necessary to turn it on manually (using the green button on the top edge). 

At this point a Palm logo will appear on the screen and the unit will reboot (i.e. the operating system will 
reload) as indicated by moving black bars on the screen.  When this is complete the user will be 
prompted to confirm or re-enter the real-time clock information (date, time, location) and confirm user 
preferences. 

When this has been completed the Palm T

|

X is ready for use, however it should be left charging for 

several more hours before disconnecting it from the ac power adaptor/charger. 

In exceptional circumstances, the Palm T

|

X may fail to re-initialise after being re-charged.  Should this 

happen it will be necessary to operate the “Reset” button on the back of the T

|

X.  To do this, remove the 

T

|

X from the PSA1301/2701 (see “Removing and Re-fitting the Palm T

|

X” above), and use the Palm 

stylus to press the reset button, see section 25.2.  It should then re-initialise. 

14.2.  Technical Note about the OFF state 

Handheld computers operate on the principle of being ‘always on’.  When they are turned ‘off’, they 
merely go into a low power consumption mode with the display and other functions disabled - this is 
often referred to as Sleep mode. When turned back on, they restart instantly without the delay that would 
be caused if the operating system had to be reloaded. 

Consequently when ‘off’, a small amount of power is still being taken from the battery.  It is assumed that 
the user will keep the unit charged and never let the battery fully discharge.   

In most handheld computers, allowing the battery to fully discharge would cause all of the user’s data to 
be lost.  However, the Palm T

|

X uses permanent Flash memory which prevents user data from being lost 

if the battery is fully discharged.  The real-time clock, however, cannot continue to run with a fully 
discharged battery. 

If the unit is left unused and unconnected to its charger for an extended period (typically several weeks) 
the battery will discharge to a point where re-initialisation is needed. 

The batteries of the PSA1301/2701 will also discharge over an extended period of time, but no re-
initialisation should be required. 

15.

 

 Language Choice for the Palm T|X 

As supplied with the PSA1301/2701T, the operating language of the Palm T|X has been set to English.  
It is possible to change this to French, German or Spanish, but the change should be made before the 
product is first used.  Note that the Spectrum Analyzer software remains in English. 

See the Read This First leaflet for instruction on changing the language. 

Содержание PSA1301T

Страница 1: ...THURLBY THANDAR INSTRUMENTS INSTRUCTION MANUAL PSA T Series PSA1301T PSA2701T 1 3GHz 2 7GHz RF Spectrum Analyzers...

Страница 2: ...ency Measurement 8 6 2 Amplitude Measurement 8 6 3 Sweep 9 6 4 Signal Input 10 6 5 Demodulation Zero Span mode 10 6 6 Display 10 6 7 Memory Storage 10 6 8 Connectors 11 6 9 Power Sources 11 6 10 Mecha...

Страница 3: ...2 18 2 Selecting and Leaving the Application 22 18 3 Touch Screen or Hard Key operation 22 18 4 The PSAnalyzer Screen 23 18 5 Operation and Navigation 23 18 6 Turning Off Power Saving 24 19 The PSAnal...

Страница 4: ...sing a printer connected to a PC 40 22 3 Transferring PSAnalyzer Files to a Personal Computer 40 23 PSAnalyzer Files and Defaults 41 23 1 Factory Default Settings for PSAnalyzer 41 23 2 File Locations...

Страница 5: ...ch are shown as follows see section X X Within a PDF file the boxed number is a hyperlink to that section number thus enabling the user to jump rapidly the section referred to and then jump back to co...

Страница 6: ...primary function of RF spectrum analysis without linkage to other devices TTi provides a USB linked memory card reader in order that PSAnalyzer files can be transferred to a PC where this is required...

Страница 7: ...d key Some general information regarding the capabilities of the unit is available from the Info key on the Exit menu 4 2 Detailed Operating Guide To fully understand the operation of the PSA1301 2701...

Страница 8: ...cally peak finding and tracking A reference trace can be displayed in addition to the live trace Limit lines can also be created and displayed The graticule is fully controllable Any number of traces...

Страница 9: ...ppm over 10o C to 30o C Ageing Better than 3 ppm per year Phase Noise Phase Noise Typically 100dBc Hz at 100kHz offset at 500MHz PSA1301 Typically 90dBc Hz at 100kHz offset at 500MHz PSA2701 Resoluti...

Страница 10: ...age 55dBc typically 60dBc Residual Spurii Below noise floor PSA1301 3dB above noise floor PSA2701 Amplitude Markers Number of Markers One Two or None Marker Resolution 0 1 dB Amplitude Readout Display...

Страница 11: ...Stored trace either recalled from memory or copied directly from live trace Selectable on or off Resolution Modes Selectable as High Resolution or Low Resolution In low resolution mode the trace area...

Страница 12: ...from mains power using the AC power adaptor provided by TTi This powers and recharges both the Spectrum Analyzer and the Palm T X simultaneously AC Adaptor Charger Input Voltage Range 100V to 240V no...

Страница 13: ...use which angles the unit at approximately 25 degrees to the horizontal and can alternatively be used as a hook mount 6 11 Environmental and Safety Operating Range 5 o C to 40 o C 20 to 80 RH Storage...

Страница 14: ...C Compliance was demonstrated by conformance to the following specifications which have been listed in the Official Journal of the European Communities EMC Emissions a EN61326 2006 Radiated Class B Im...

Страница 15: ...e susceptibility to high electromagnetic fields throughout their measurement range In general the instrument will operate correctly but may show the interfering frequencies on the measurement trace if...

Страница 16: ...that the shell of the mini USB connector is galvanically connected to the body of the SMA input which will therefore be at earth ground potential when the USB port is connected to a desktop PC To main...

Страница 17: ...echarging When they are fully charged this LED will go off The charging state and charge level of the Palm T X batteries is shown on the display of the T X see section 17 1 10 3 Demodulated Audio Outp...

Страница 18: ...available The memory card can be used as a means of transferring data to and from a PC using the USB card reader supplied See section 22 3 for details 12 1 Removing and Inserting the Card The card fi...

Страница 19: ...Removing Re fitting Remove the Stylus from the Palm T X Push the Palm T X onto the connector by pushing it firmly in the line of the instrument Ensure that the two fixing latches are open Locate the t...

Страница 20: ...tialise 14 2 Technical Note about the OFF state Handheld computers operate on the principle of being always on When they are turned off they merely go into a low power consumption mode with the displa...

Страница 21: ...Pressing it again turns it back on Note It may be necessary to press the key more than once to turn on because of slow key polling in sleep mode The PSAnalyzer application suppresses automatic switch...

Страница 22: ...hours of continuous operation from a charge Recharge time is typically less than four hours To conserve the battery the PSAnalyzer application can be put into sleep mode when not in use by pressing a...

Страница 23: ...r and turn the function off The USB lead should be unplugged at the spectrum analyzer end The USB cable could induce interference into the spectrum analyzer that could cause spurious signals to be dis...

Страница 24: ...Screen Data Area 2 Icon Area Data Area 1 Primary Frequency Settings Reference Level Resolution B W Video Filter Sweep Run Stop Battery Status Data Area 3 Data Area 4 Marker Readout Sweep Type Attenuat...

Страница 25: ...o Span or Quick Menu Also provides a function called Preset which sets defined conditions from a single key press Note that the labels of the next two keys depend upon the mode setting Centre set the...

Страница 26: ...jacent to the RF Signal Input Demodulation mode and audio level are selected from the AM FM menu Quick Menu is a different mode of operation which provides direct access to some of the most commonly u...

Страница 27: ...and return to the home menu press the Back Home hard key or the grey area at the top left of the menu 19 5 Step Value Sub Menu There are four ways to set the step value Set Step M delta sets the step...

Страница 28: ...y area at the top left of the menu 19 8 Stop Frequency Menu There are three ways to set the stop frequency Stop M1 sets the stop frequency equal to the frequency at the position of Marker One Start M1...

Страница 29: ...B range To exit this menu and return to the home menu press the Back Home hard key or the grey area at the top left of the menu 19 11 Markers Menu Turn Markers On press the centre of the 5 way navigat...

Страница 30: ...hs of 280 kHz and 15 kHz PSA1301 or between the three resolution bandwidths of 1 MHz 280 kHz and 15 kHz PSA2701 Normal sets a continuous sweep mode where a new sweep is commenced automatically after t...

Страница 31: ...limit lines have been created or recalled the key will have no function The colour of Limit Line 1 is red Limit line 2 is blue To exit this menu and return to the home menu press the Back Home hard k...

Страница 32: ...ered to create limit lines for the display and to save and recall limit lines from memory see section 20 9 To exit this menu and return to the home menu press the Back Home key or the grey area at the...

Страница 33: ...Marker control and no longer operate the key highlighting The purpose of Quick Menu is to provide direct access to some of the more commonly used functions without having to select sub menus Frequenc...

Страница 34: ...enus see section 19 Moving between menus requires a return to the Home menu by pressing the Back Home hard key When using the touch screen for control rather than the 5 way navigator it is also possib...

Страница 35: ...equired this can be used to edit the number after it is entered and before it is confirmed using standard Palm cursor based editing See section 20 1 for details of how to change the input area type Pr...

Страница 36: ...sweep and the old sweep values are sequentially over written with the new values In this mode the progress of the sweep is indicated by a moving yellow line at the bottom of the graticule Video Filter...

Страница 37: ...s described in section 19 11 Markers are vertical lines with a box that highlights the intersection point of the marker with the trace The readout shows the frequency and level at the intersection poi...

Страница 38: ...o store traces and complete instrument set ups to a memory card and recall them as required See Stores menu section 19 14 Each key within the Stores menu brings up a dialogue box which includes a scro...

Страница 39: ...is deletes all points in the current list To exit the editor screen while saving all of the changes press the Finish key To exit without saving any changes press the Cancel key Note that frequency amp...

Страница 40: ...r measurement accuracy will be achieved using Hi Res Mode 20 11 Changing the Palm Input Area type The Set by Keyboard function for frequency setting and all Store Recall functions brings up the standa...

Страница 41: ...anywhere on the screen will return to the thumbnails view Press the home key to exit the Media application 22 2 Printing a Stored Screen Image using a printer connected to a PC Screen image files are...

Страница 42: ...pen the disk drive under its allocated drive either from My Computer or from Explorer The directory structure of the card will be visible Trace files are stored in the folder PSAnalyzer Tr21aces Set u...

Страница 43: ...s 120 kilobytes and is an image of 379 pixels high by 320 pixels wide in 8 bit indexed colour format When transferred to a personal computer screen files can be opened in any image editing program and...

Страница 44: ...4 4 Calibration To ensure that the accuracy of the instrument remains within specification the calibration should be checked and if necessary adjusted annually The calibration adjustment procedure is...

Страница 45: ...er button 2 While holding the power button use the tip of the stylus or a similar object without a sharp tip to gently press and release the reset button 3 Wait for the Palm Powered logo to appear not...

Страница 46: ...Shortcut to PSAnalyzer As supplied by TTi the PSAnalyzer application is installed within the Palm T X applications so as to occupy the first position within the Applications screen category All or Unf...

Страница 47: ...idgeshire PE29 7DR England United Kingdom Telephone 44 0 1480 412451 Fax 44 0 1480 450409 International web site www tti test com UK web site www tti co uk Email info tti test com Aim Instruments and...

Отзывы: