11
Store Screen:
This function copies the whole screen area to memory as a bit-map. Any
number of screens can be stored under either default file names or user
entered file names. Screen bit-maps can be viewed using the “Media”
application of the Palm T
|
X, or transferred to a PC for printing.
Store/Recall Limit
Lines
Any number of limit lines can be stored under either default file names or user
entered file names, and recalled as required.
6.8. Connectors
RF Input Connector:
Standard SMA connector.
Power Connector:
1.3 mm dc power socket (centre positive) for 5.2V/1A external AC power
adaptor/charger as supplied by TTi.
USB Connector:
Mini USB connector (5 pin mini B) which provides direct access to the USB port
of the Palm T
|
X.
Audio Connector:
3.5 mm jack socket for demodulated audio out (accepts mono or stereo plugs).
6.9. Power Sources
AC Line Operation/Charging
The PSA1301/2701T can be operated from mains power using the AC power adaptor provided by TTi.
This powers and recharges both the Spectrum Analyzer and the Palm T
|
X simultaneously.
AC Adaptor/Charger
Input Voltage Range:
100V to 240V nominal 50Hz/60Hz. Interchangeable plus for UK, Euro, USA
and Australia are supplied.
Battery Operation
The PSA1301/2701 and the Palm T
|
X contain their own internal rechargeable battery packs. The
PSA1301/2701 is turned off automatically whenever the PSAnalyzer program is exited, or when the
Palm T
|
X is turned off.
PSA1301/2701 Spectrum Analyzer
Battery Type:
Ni-Mh 3.6V 700mA-hr (3 x AAA cell)
Battery Life:
> 5 hours continuous operation (PSA1301)
> 4 hours continuous operation (PSA2701)
Recharge Time:
< 3 hours from fully discharged
Palm T|X
Battery Type:
Li-ion 3V 1050mA-hr
Battery Life:
Typically 4 hours continuous operation (depending on backlight brightness)
Recharge Time:
< 4 hours from fully discharged
Auto Sleep Mode
To conserve battery life, the system can be set to automatically switch into sleep mode after a defined
time from the last key press. This can be set between 5 mins and 60 mins (or never).