25
Explanation of command
4.2.36
DATA?
(lump read-out of test result)
Function
Reads out the detail data of test result.
[Command to use after the finish of the test (READY status)]
Judgement result and data are retained until the next start even if the
RESET command is made or
STOP
switch is pressed.
Structure
DATA?
Transmission
DATA?
Response
[Example of response after the finish of single AC withstanding voltage test]
Judgement result and action
during the test
Response
AC W test passed
JUDGE=GOOD,AJUDGE=GOOD,VOLT=1.51kV,CURRENT=1.23mA
AC W test failed for HIGH
JUDGE=NG,AJUDGE=HIGH,VOLT=1.51kV,CURRENT=32.1mA
AC W test failed for LOW
JUDGE=NG,AJUDGE=LOW,VOLT=1.51kV,CURRENT=0.15mA
At RESET (Stop)
Note-1
JUDGE=NULL,AJUDGE=NULL,VOLT=0.00kV,CURRENT=0.0mA
At occurrence of PROTECT
Note-2
JUDGE=PROTECT,AJUDGE=HIGH LOW,
VOLT=1.50kV,CURRENT=1.23mA
[Example of response after the finish of single DC withstanding voltage test]
Judgement result and action
during the test
Response
DC W test passed
JUDGE=GOOD,DJUDGE=GOOD,VOLT=1.51kV,CURRENT=1.23mA
DC W test failed for HIGH
JUDGE=NG,DJUDGE=HIGH,VOLT=1.51kV,CURRENT=10.0mA
DC W test failed for LOW
JUDGE=NG,DJUDGE=LOW,VOLT=1.51kV,CURRENT=0.15mA
At RESET (Stop)
Note-1
JUDGE=NULL,DJUDGE=NULL,VOLT=0.00kV,CURRENT=0.0mA
At occurrence of PROTECT
Note-2
JUDGE=PROTECT,DJUDGE=HIGH LOW,
VOLT=1.50kV,CURRENT=1.23mA
Note-1
: Data is 0.
Note-2
: Responses with the data at the occurrence of PROTECT.
For the test which could not be performed, the data is 0.