12
7. Setting of test mode
7.1
●
Status of display and expression in instruction manual
Digital display
Flat display
LED lamp
Lit-up mode
GOOD
●
KEY LOCK
Blinking mode
GOOD
◎
KEY LOCK
Turn-off mode
GOOD
○
KEY LOCK
7.2
●
Selection of each test mode
The following 2 test modes can be set.
AC
(AC Withstanding voltage test)
DC
(DC Withstanding voltage test)
①
To enter the setting for selection of test mode,
In READY status, press
or
key, then the test mode lamp blinks.
The test mode lamp moves up and down with
▲
or
▼
key. Set to the required
test mode lamp (blinking status). A press of
ENTER
key determines the selected
mode and the tester returns to
READY
status
②
To enter the setting of AC withstanding voltage test,
In the condition
①
, while
AC
(AC withstanding voltage test) is in blinking, press
or
key, then you can enter the setting of test condition for the AC
withstanding voltage test.
③
To enter the setting of DC withstanding voltage test,
In the condition
①
, while
DC
(DC withstanding voltage test) is in blinking, press
or
key, then you can enter the setting of test condition for the DC
withstanding voltage test.
PROTECT
○
ONLINE
○
○
○
○
○
KEY LOCK
DISCHARGE
REMOTE
READY
AC
DC
○
No.
s
mA
SET
SET
LOW
GOOD
LOW
MODE
JUDGE
CURRENT
TEST VOLTAGE
MEMORY
RANGE
TIME
kV
kV
●
AC
DC
HIGH
HIGH