16
Setting of test condition for withstanding voltage test
8.3
●
High limit of leak current
Adjustable range:
For AC withstanding voltage test, 0.1~110.0mA.
For DC withstanding voltage test, 0.1~11.0mA.
Note:
The high limit value of leak current can not be set lower than that of low limit, so
please set to comply with the following conditions.
1. When the low limit value of leak current is determined, set the high limit value to
exceed the value of low limit.
2. When the high limit value of leak current is determined, set the low limit value
not to exceed the value of high limit, or turn “OFF” the low limit.
To enter setting mode
①
In READY status, press
or
key, then the test mode lamp blinks. The test
mode lamp moves up and down with
▲
or
▼
key. Make the required test
mode lamp blinking (
AC
or
DC
).
②
Press
or
key and make the test mode lamp lit up.
Setting of high limit of leak current
①
Press
or
key and select the status that the current display blinks and
HIGH SET
is lit up (refer to the above figure).
②
Next, press
▲
or
▼
key and set the high limit of leak current value.
A press of
key (
SHIFT
and
▲
keys at a time) or
key (
SHIFT
and
▼
keys at a time) allows the setting of second digit.
To return to the previous setting
Press
key, then the setting changes to the
setting of referential voltage
To the next setting
Press
key, then the setting moves to the
setting of low limit of leak current
.
Finish of setting
Press
ENTER
key, then the tester returns to READY status, memorizing the settings
having been made.
When the
EXIT
key (
SHIFT
and
ENTER
key at a time) is pressed in the setting
mode, the setting mode for the test condition is interrupted and the tester becomes
READY
status.
The test conditions in this case are those before entering the setting mode of test
condition.