40
20. Specifications
1. Withstanding voltage test section
1.1 AC withstanding voltage output
(1) Output voltage
0~2.5kV / 0~5kV AC
(2) Output capacity
500VA (5kV, 100mA). with the rated power source voltage.
For the output current 50mA or higher, 30 min. or less continuously.
(3) Wave shape
Shape of commercial power source.
(4) Voltage fluctuation rate
15% or less
(with the rated power source voltage and at no load
⇒
max. load)
(5) Voltage output system
Zero-cross throw switch.
(6) Setting of output voltage Manual setting by volt slider.
1.2 DC withstanding voltage output
(1) Output voltage
0~2.5kV / 0~5kV DC
(2) Output capacity
50W (5kV, 10mA) at the rated power source voltage.
For the output current 6mA or higher, continuously for 1 minute or
less. (The pause time needs to be 10 times of test time.)
For the output current 1mA to less than 6mA, continuously for 2
minutes or less. (The pause time needs to be 4 times of test time.)
(3) Output ripple
At 5kV no load
: 50Vp-p, typical
At the maximum rated output : 100Vp-p, typical.
(4) Voltage fluctuation rate
3% or less
(with the rated power source voltage and at no load
⇒
max. load)
(5) Setting of output voltage Manual setting.
(6) Discharging function
Discharges the electric charge which is charged in the test sample.
2. Voltage measurement
2.1 Analog
(1) Scale
AC and DC sharing. 0~5kV
(2) Accuracy
±
5% of F.S
(3) AC indication
Effective average rectification value indication.
(4) Unit
“kV”
2.2 Digital
(1) Measuring range
AC and DC sharing. 0.00~6.00kV
(2) Display
Digital display in 3 digits, green LED, character height 10mm.
(3) Accuracy
±
1.
5% of F.S (F.S 2.5kV/5kV)
(4) Voltage display
During the test, the voltage applied to the high voltage terminal is
displayed.
At the finish of the test, the voltage value at judgement is retained.
At READY, the referential voltage value is displayed.
(5) AC indication
Effective average rectification value indication
3. Current measurement
(1) Display range
AC:0.01~199.9mA (2 ranges, joint change-over with high limit value)
DC:0.01~19.99mA
(2) Display
Digital display in 3 1/2 digits, green LED, character height 10mm.
(3) Resolution
AC:0.01mA (0.1~9.9mA)
0.1mA (10.0~110.0mA)
DC:0.01mA (0.1~11.0mA)
Note
: ( ) shows high limit set value.
(4) Accuracy
±
(5%+20
μ
A) of high limit set value.
(5) Current display
During the test, the leak current value is displayed.
At the finish of the test, the leak current value at judgement is retained.
At READY, the high limit value is displayed.
(6) AC indication
Effective average rectification value indication.