41
Specifications
4. Judgement of test result
(1) Judgement system
High limit
Analog comparator.( For short-circuit
detection, set value internally fixed. )
High and low limit
Digital comparator.
(2) Adjustable range
High limit AC:0.1~110.0mA (low limit + 1 digit or more)
DC:0.1~11.0mA (low limit + 1 digit or more)
Resolution 0.1mA (0.1~110.0mA, for DC up to 11.0mA)
Low limit
AC:0.0~109.0mA (high limit – 1 digit or less)
DC:0.0~10.9mA (high limit – 1 digit or less)
Resolution 0.1mA (0.0~109.0mA, for DC up to 10.9mA)
Note1
: Low limit setting can be ON/OFF
Note2
: High and low limit value for the AC withstanding voltage test
and the same for the DC withstanding voltage test are
memorized as independent item with each other.
(3) Judgement condition
High limit value
>
Leak current
>
Low limit value … GOOD
High limit value
≦
Leak current ………………………. HIGH NG
Low limit value
≧
Leak current ………………………. LOW NG
Note:
Output time of GOOD judgement can be switched to
continuous or 0.2s.
For the AC withstanding voltage testers, the leak current due to the
capacity distribution in the high voltage cable, jig and so on can
cause the judgement error.
Please determine the judgement criterion value, taking this leak
current into account.
The following values are the referential values on condition that the
wiring is made, keeping the distance between HIGH voltage side
cable (red) and LOW voltage side cable(black) of the attached high
voltage cable (5880-25-020).
Output voltage
1kV
2kV
3kV
4kV
5kV
Leak current
10
μ
A
20
μ
A
30
μ
A
37
μ
A
47
μ
A
5. Test time
(1) Adjustable range
0.5~999s, with time off function.
(2) Setting resolution
0.1s (0.5~99.9s) / 1s (100~999s)
(3) Time display
0.0~999, 3 digits green LED, character height 8mm
During the test
With timer ON
Remaining time is displayed.
With timer OFF Time lapse is displayed.
At READY, set value is displayed.
(4) Accuracy
±
20ms (0.5~99.9s) /
±
200ms (100~999s)
6. Input/output signal
(1) Connector
36P Anphenol connector on the rear panel.
(2) Output signal
Open collector 30V DC, 400mA MAX (TOTAL)
(3) Name of output signal
TEST
: In test.
AC-TEST
: In AC withstanding voltage test.
DC-TEST
: In DC withstanding voltage test.
END
: Finish.
TEST/H.V. OUT : While high voltage is output at output terminal.
READY
: In waiting.
GOOD
: At good judgement (0.2s/continuous changeable).
NG
: At NG judgement (continuous).
HIGH NG
: At NG judgement for high limit (continuous).
LOW NG
: At NG judgement for low limit (continuous).
PROTECTION : While protective function is activated.
(4) Power source for
output/input signal
24V DC, 0.1A
(5) Input signal
H=16.8~24V, L=0~3.8V
1c=10mA, L level minimum pulse width=40ms