32
External control
Connector used: 36P Anphenol
Note
: When externally remote controlled, REAR:ON and COM are short-circuited.
The operation is same as that of
REMOTE
terminal, the article 13.2 (P28).
14.3
●
Interlock signal
The interlock is the function to shut off the output getting the tester to jointly work with
the external device, in order to secure the safety of operator.
By making open the pin 5 (INTER-LOCK) of the
REMOTE/OUT
connector
⑱
on the
rear panel, the tester becomes interlock status and the start of the test is disabled.
During the interlock function is in operation,
is displayed, the output of
8526
is shut off and the operation of all the switches are disabled.
To cancel the interlock, short-circuit the pin 5 and pin 23 (COM) of the
REMOTE/OUT
connector
⑱
to make it to “L” level, and then press the
STOP
switch
②
.
is turned off and
READY
lamp is lit up, enabling the test.
Note
: The pin 5 and 23 of the attached REMOTE/OUT plug (36P) are short-circuited.
Considering the safety aspect, please provide a proper interlock solution to jointly work
with the external device, for example, as the following connection example shows.
REMOTE/OUT connector
External device
Safety security switch on test jig
Fig.14.1
Interlock connection example
14.4
●
Protective function (PROTECTION)
The protective function is the action that the PROTECTION is output from the
REMOTE/OUT
connector
⑱
on the following condition.
●
When the discharging of the test sample does not finish even after passing 10 seconds
from the finish of test.
●
When the voltage output does not fall even after passing 10 seconds from the finish of
test.
●
When the interlock input is turned OFF.
●
When the remote status is changed during the test.
●
When the test voltage becomes out of the range of referential voltage and the test is
stopped.
●
When there is an abnormality in the DC high voltage power source circuit.
1
18
19
36
18
36
1
19
5
23