Specifications
53
Maintenance:
Factory clean/calibrate ..... Recommended annually
User zero calibration ........ Before each use
User flow calibration ......... As needed
Communications Interface:
Type ................................. USB 1.1
Connector, instrument...... USB Mini-B (socket)
Minimum Computer Requirements for TrakPro Software:
Communications port ....... Universal Serial Bus (USB) v1.1 or higher
Operating system ............. Microsoft Windows
®
XP (32-bit) or
Windows
®
7 (32-bit or 64-bit) operating
systems
Approvals:
Intrinsic Safety
Rating Information ............ TSI Battery Pack P/Ns 801722, 801724,
801728, 801729 or 801743
INTRINSICALLY SAFE CSA
Exia
T2A with 801724, 801729 or 801743
T2C with 801722, 801728
Class I Groups A, B, C, D
Class II Groups E, F, G
Class III
File: 200507
0344
Immunity EN61326-1:1997 + A11998 Clause 6
Emissions EN61326:1997 + Amendment A1:1998
!
C a u t i o n
All Dust Plugs and gaskets must be installed on the
SidePak AM510, in addition to using the above battery
packs to achieve Intrinsic Safety Rating.
Содержание SIDEPAK AM510
Страница 1: ...SIDEPAK PERSONAL AEROSOL MONITOR MODEL AM510 USER GUIDE ...
Страница 2: ......
Страница 66: ...Appendix B 58 This page intentionally left blank ...
Страница 69: ...61 Appendix D CSA Certificate of Compliance ...
Страница 70: ...Appendix D 62 ...
Страница 71: ...CSA Certificate of Compliance 63 ...
Страница 72: ...Appendix D 64 This page intentionally left blank ...
Страница 73: ...65 Appendix E AM510 Quick Reference Guide ...