Troubleshooting 61
Symptom
Possible Causes
Solution
Suspicious
readings
(If the
P
ORTA
C
OUNT
passes the Daily
Checks, the
problem is the
respirator, not the
P
ORTA
C
OUNT
.)
P
ORTA
C
OUNT
®
P
RO
fit
tester is flooded with
alcohol.
Tube fittings on
P
ORTA
C
OUNT
®
P
RO
are
loose.
P
ORTA
C
OUNT
®
P
RO
leaks.
Run 2 hours to dry the
optics (remove cartridge
and install storage cap).
Tighten with pliers.
Zero-Check the
P
ORTA
C
OUNT
®
P
RO
fit
tester and fix any leaks.
Sample lines too long.
Use standard 5-foot Twin
Tube Assembly.
Alcohol Cartridge is
loose.
Tightly close the Alcohol
Cartridge.
Respirator is not
equipped with HEPA
filters.
Install HEPA filters for fit
tests.
Twin Tube Assembly is
kinked, pinched, or
blocked.
Straighten out the Twin
Tube Assembly or
remove the obstruction.
Leaking
respirator
probe.
Tighten or seal probe.
Tubing in fit test adapter
is kinked or pinched.
Straighten out tubing.
Respirator is faulty.
Fix or replace the
respirator.
Particle count is
zero or near zero
Alcohol level is low.
Replenish alcohol.
Sampling through HEPA
Filter.
Remove filter.
Twin Tube Assembly is
blocked.
Remove the blockage.
Moisture build-up in
Alcohol Wick.
Change Alcohol Wick
inside Alcohol Cartridge.
Run 2 hours to dry optics
(remove the alcohol
cartridge and install
storage cap).
Содержание portacount pro 8030
Страница 2: ......
Страница 10: ......
Страница 22: ......
Страница 30: ......
Страница 61: ...Service and Maintenance 55 ...
Страница 62: ......
Страница 72: ......
Страница 76: ......
Страница 80: ......
Страница 82: ......
Страница 86: ......
Страница 87: ...81 Appendix G The Rainbow Passage ...
Страница 88: ......