52 Chapter
6
3.
Examine the white Alcohol Wick. Discard wicks that are severely
discolored or physically damaged. Some light brownish-yellow
discoloration of the wick is normal.
If the wick is in good condition, dry it by placing it on a clean surface in
a well-ventilated area and allow it to dry for at least two days.
4.
Before re-installing the alcohol wick, make certain that all parts are
clean. Small bits of the wick or lint can cause serious problems if they
get into the P
ORTA
C
OUNT
®
P
RO
fit tester.
5.
Inspect the inside surfaces of the Alcohol Cartridge and the Wick
Retainer Cap. Blow air into them, if necessary, to make certain that
there are no dust particles present.
6.
Blow air onto all surfaces of the Alcohol Wick that will be used to make
certain that any loose particles that may have shed from the wick are
removed.
7.
Examine both ends of the wick. You will notice that one end is
smoother than the other. Slide the smooth end of the wick into the Wick
Retainer Cap and push firmly until the wick hits bottom.
8.
Blow everything off again.
9.
Align the two halves of the Alcohol Cartridge and press firmly until
they snap together.