49
Chapter 6
Service and Maintenance
Recalibration Interval
TSI recommends that the P
ORTA
C
OUNT
®
P
RO
Respirator Fit Tester be
cleaned and recalibrated annually. Try to arrange your fit testing schedule to
allow for annual factory recalibration. If your fit testing schedule is seasonal
or you expect heavy usage during an upcoming period (such as an outage),
TSI recommends that you have the P
ORTA
C
OUNT
®
P
RO
fit tester serviced
prior to that time. TSI's Customer Service Department provides fast service
in order to minimize your down time. Contact TSI on the Internet at
25H
www.tsi.com
or send e-mail to
26H
.
Status Messages
Two status messages can be displayed by the P
ORTA
C
OUNT
®
P
RO
Respirator
Fit Tester.
Low Particle Message
The P
ORTA
C
OUNT
®
P
RO
fit tester is factory
programmed not to allow fit testing when the
ambient particle concentration is below 1000
particles/cm
3
for the Model 8030 or 8038 when
testing masks with % efficiencies or greater and
30 particles/cm
3
for the Model 8038 when
testing masks with <99% efficiencies. If the
ambient sample in Fit Test Mode is below these
levels, the “Low Particle” message appears on
the touch screen display and the fit test is
automatically terminated.
This message appears in Fit Test Mode only, for
one of the reasons described below. Consult the
27H
Troubleshooting
chapter if necessary.
Low Alcohol Level
Message
When the "Low Alcohol Level" message is
displayed, it means that the P
ORTA
C
OUNT
®
P
RO
fit tester is low on alcohol. This message does not
necessarily mean that the P
ORTA
C
OUNT
®
P
RO
fit
tester will stop working soon. Fit test results are
accurate even when this message is on. Refer to
the
28H
Troubleshooting
chapter for other possible
causes and solutions for this message.
Содержание portacount pro 8030
Страница 2: ......
Страница 10: ......
Страница 22: ......
Страница 30: ......
Страница 61: ...Service and Maintenance 55 ...
Страница 62: ......
Страница 72: ......
Страница 76: ......
Страница 80: ......
Страница 82: ......
Страница 86: ......
Страница 87: ...81 Appendix G The Rainbow Passage ...
Страница 88: ......