54 Chapter
6
Reinsert the alcohol cartridge. Turn the P
ORTA
C
OUNT
®
P
RO
Respirator Fit
Tester on. Perform the Daily Checks. If the problem persists, try
Procedure B.
Procedure B
Use this procedure only if Procedure A fails to clear the internal nozzle.
Turn the P
ORTA
C
OUNT
®
P
RO
fit tester off and remove the alcohol
cartridge.
Turn and hold the P
ORTA
C
OUNT
®
P
RO
fit tester up on end so that you
can look down into the cartridge cavity.
Take a bottle of the alcohol used to operate the P
ORTA
C
OUNT
®
P
RO
fit
tester and drip 3 to 4 drops down into the cartridge cavity. Try to keep
the drops centered so they hit the bottom without touching the sides. Do
not use too much. One drop is enough if it all reaches the nozzle.
Keep the P
ORTA
C
OUNT
®
P
RO
fit tester in this vertical position for about
5 minutes so the alcohol can soften or dissolve the blockage.
Put the P
ORTA
C
OUNT
®
P
RO
Respirator Fit Tester down in normal operating
position and perform Procedure A.
Содержание portacount pro 8030
Страница 2: ......
Страница 10: ......
Страница 22: ......
Страница 30: ......
Страница 61: ...Service and Maintenance 55 ...
Страница 62: ......
Страница 72: ......
Страница 76: ......
Страница 80: ......
Страница 82: ......
Страница 86: ......
Страница 87: ...81 Appendix G The Rainbow Passage ...
Страница 88: ......