Conducting a Fit Test with the P
ORTA
C
OUNT
®
P
RO
Respirator Fit Tester
35
b.
If you want to select a different respirator, click the arrow on the
Respirator List
field to display a list of all the respirators in the
database. Use the stylus to select the respirator you want to use,
click
Next
to select the current protocol.
Note:
You cannot add respirator records to the database here. If
the respirator you want to use for fit tests is not in the
database, or if no records are in the database, you must
exit the Fit Test function and use the Database tab to
access the respirator database (see
18H
Chapter 5
). You can
also use F
IT
P
RO
™
software to update the database to
include the respirator.
6.
After you press
Next
, the
Current Protocol
dialog box appears, the
mask size and operator fields are blank. You must fill in these fields
before you can continue.
7.
Make certain the exercise protocol shown in the Current Protocol field
is the one you want to use. If you want to select a different respirator,
click the arrow on the
Current Protocol
field to display a list of all the
Protocols in the database. Use the stylus to select the respirator you
want to use.
Note:
You cannot add a protocol to the database here. If the protocol
you want to use for fit tests is not in the database, or if no
records are in the database, you must exit the Fit Test function
and use the Database tab to access the protocol database (see
19H
Chapter 5
). You can also use F
IT
P
RO
™
software to update the
database to include the respirator.
8.
Click on the
Mask Size
field and use the on-screen keyboard to type in
the mask size, for example: SMALL, MEDIUM, or LARGE. You can
also use abbreviations such as S, M, or L.
9.
Click on the
Operator
field and fill in your name or initials.
10.
If you want to change the
Due Date
for this person’s next fit test, click
on the down arrow and select a different date from the calendar.
11.
When you have filled in this dialog, click
Next
to proceed.
12.
After you click
Next
, the
Run Test: Fit Test Step 4 of 4
dialog
appears, and you are ready to begin the fit test. Before you click
START
to begin the test, verify the information at the top left of the
screen and make certain the
P
ORTA
C
OUNT
®
P
RO
sample tube is
connected to the respirator (Use one of the Tube Adapters if necessary.),
and ask the test subject if they are ready to begin the exercise protocol
immediately.
Содержание portacount pro 8030
Страница 2: ......
Страница 10: ......
Страница 22: ......
Страница 30: ......
Страница 61: ...Service and Maintenance 55 ...
Страница 62: ......
Страница 72: ......
Страница 76: ......
Страница 80: ......
Страница 82: ......
Страница 86: ......
Страница 87: ...81 Appendix G The Rainbow Passage ...
Страница 88: ......