58 Chapter
7
Symptom
Possible Causes
Solution
Low Alcohol
Level message
Low on alcohol.
Refill Alcohol Cartridge.
Excessive
moisture
inside P
ORTA
C
OUNT
®
P
RO
fit tester.
Change Alcohol Wick
inside Alcohol Cartridge.
Run 2 hours to dry optics
(remove the alcohol
cartridge and install
storage cap).Dump old
alcohol from the Alcohol
Fill Capsule and add fresh
alcohol.
P
ORTA
C
OUNT
®
P
RO
fit
tester
needs recalibration
and cleaning.
Return to factory for
service.
Nozzle is plugged.
See
33H
Nozzle Cleaning
in
Service and Maintenance
chapter.
Low Particle
message
Low on alcohol.
Replenish alcohol.
Hoses are reversed on
Twin Tube Assembly.
Verify the hoses are
connected properly to the
P
ORTA
C
OUNT
®
P
RO
and
mask.
Wrong tube is connected
to respirator.
Connect the proper tube
to the respirator (clear
tube).
Twin Tube Assembly is
kinked, pinched, or
blocked.
Straighten out the Twin
Tube Assembly or
remove the obstruction.
Particle count in area
really is low.
Move to another location
or use a particle generator
(such as a candle or an
ultrasonic humidifier with
tap water).
Содержание portacount pro 8030
Страница 2: ......
Страница 10: ......
Страница 22: ......
Страница 30: ......
Страница 61: ...Service and Maintenance 55 ...
Страница 62: ......
Страница 72: ......
Страница 76: ......
Страница 80: ......
Страница 82: ......
Страница 86: ......
Страница 87: ...81 Appendix G The Rainbow Passage ...
Страница 88: ......