3000A SERIES CALIBRATION MANUAL
TRANSMILLE LTD. Version 3 : Jun 2014
Page 82
3. The Calibrator will now change to the A/D Input Screen. Use the Range Up /
Range Down keys to select the Zero Calibration adjustment point
4. Apply an output of 0V from your voltage source to the adapter interface, using Pin
7 as the Ground or Negative connection, and Pin 9 as the Signal or Positive
connection.
5. Wait for the reading to stabilise on the screen of the 3000A. After achieving a
stable reading, press the ‘Store’ button.
7. The following 2 screens are displayed briefly to confirm that the calibration has
been saved.