3000A SERIES CALIBRATION MANUAL
TRANSMILLE LTD. Version 3 : Jun 2014
Page 17
Calibration of Ranges
D.C. Voltage
To calibrate the D.C. Voltage parameter, complete the procedure as follows:
1.
Select
CALIBRATE
using the soft key
2.
Use the ‘Digital Control’ or the ‘Arrow Keys’ to highlight the required
parameter ‘DC Volts’ and then press
SELECT
soft key
3.
Connect the calibrator output voltage terminals to the precision Multimeter.
Ensure that the Multimeter has been zeroed as a system by shorting out the
leads and pressing the null button.
4.
Use the ‘range up’ and ‘range down’ keys to change the calibration point to
‘Zero Calibration’.