3000A SERIES CALIBRATION MANUAL
TRANSMILLE LTD. Version 3 : Jun 2014
Page 77
3. The Calibrator will now change to the 200uA DC range. Use the Range Up /
Range Down keys to select the Positive Full Scale adjustment point
4. Connect the calibrator output voltage terminals to the precision multimeter. Ensure
that the multimeter is in DC Current on the appropriate range. Ensure that the correct
terminals / current shunt is used for the various current outputs. The 3000A
calibrators can output up to 30A AC current, which can blow the fuse / cause
damage if the incorrect input is used
5. Press and measure the output.
6. To adjust the output enter the value as measured on the multimeter
E.g. measured output = 99.90 uA, use the function control keys
5. The output can also be adjusted by moving the ‘cursor’ to the required digit and
adjusting the output using the up / down arrows or the digital control
9
9
•
9
0
ENTER
OUTPUT
ON