3000A SERIES CALIBRATION MANUAL
TRANSMILLE LTD. Version 3 : Jun 2014
Page 50
3. The Calibrator will now change to the 1uF Capacitance range.
4. Connect the calibrator Voltage (Capacitance) terminals to the LCR Bridge. To
ensure an accurate measurement connect as a 4 wire measurement, with both
positive leads connected together. Ensure that the LCR Bridge is nulled before
connecting to the Calibrator
5. Press and measure the output.
6. To adjust the output enter the value as measured on the multimeter
E.g. measured output = 1.0016 uF, use the function control keys
5. The output can also be adjusted by moving the ‘cursor’ to the required digit and
adjusting the output using the up / down arrows or the digital control
OUTPUT
ON
1
•
0
0
1
6
ENTER