3. The Calibrator will now change to the 10mV/Division Oscilloscope Amplitude
range.
4. Connect the calibrator Oscilloscope output to the precision digital multimeter. Set
the 100mV DC voltage range. Normally the Oscilloscope Amplitude output is a
levelled square AC Voltage output, however in the calibration mode this is set to a
levelled DC waveform. This allows more accurate measurement than low level AC
Voltage measurements
5. Press and measure the output.
6. To adjust the output enter the value as measured on the multimeter
E.g. measured output = 59.9985 mV, use the function control keys
5. The output can also be adjusted by moving the ‘cursor’ to the required digit and
adjusting the output using the up / down arrows or the digital control