3000A SERIES CALIBRATION MANUAL
TRANSMILLE LTD. Version 3 : Jun 2014
Page 59
3. The Calibrator will now change to the 100 °C PRT range.
4. Connect the calibrator Voltage and Current terminals to the precision Multimeter
using 4 wire connection methods. If available, set the reference multimeter to
Temperature mode, otherwise use lookup tables to convert the indicated resistance
to temperature.
5. Press and measure the output.
6. To adjust the output enter the value as measured on the multimeter
E.g. measured output = 100.013 °C, use the function control keys
5. The output can also be adjusted by moving the ‘cursor’ to the required digit and
adjusting the output using the up / down arrows or the digital control
OUTPUT
ON
0
0
•
0
1
ENTER
1
3