3000A SERIES CALIBRATION MANUAL
TRANSMILLE LTD. Version 3 : Jun 2014
Page 54
Inductance
To calibrate the Capacitance parameter, complete the procedure as follows:
NOTE : The Inductance output from the 3000A series calibrator is a passive output.
This means that the value on the display is the Inductance value generated at the
terminals.
The Inductance cannot be ‘altered’ to a different output, the value stored is simply
the value of the inductor and the connections to the terminal.
1.
Select
CALIBRATE
using the soft key
2. Use the ‘Digital Control’ or the ‘Arrow Keys’ to highlight the required parameter
‘Capacit.’ and then press
SELECT
soft key
3. Use the ‘Digital Control’ or the ‘Arrow Keys’ to highlight the required range e.g.
’100mH ’ and then press
SELECT
soft key