28
The indicator uses the sampling method to determine the reference weight. This indicator does not
retain the piece weight when powered down.
1. Press the F1 key. The screen shows “100 -” which means “100%”.
2. Place the reference weight on the platform and allow the weight indication to stabilize.
3. Press the ENTER key to take the sample. If the reference weight is large enough, the
indicator now displays the relative percentage to the reference weight and the unmarked
annunciator is lit. If not, the indicator briefly displays “Add” briefly and returns back to the
“100 –“ screen.
NOTE: If you encounter the “Add” message, then the reference weight is too light for your
scale to process accurately.
4. To exit the percentage mode, press the C/CE key.
NOTE: The reference weight will NOT remain in scale memory when you exit percentage
mode.
Use the Over and Under keys to program in a % target for checkweighing.
HOLD APPLICATION (F30 = 5)
There are several specific hold modes, which are selected thru the F35 setting:
AUTOMATIC (F35 = 1) – Automatically locks weight on the display when stable. The percentage
hold setting (F36) unlocks the held reading and relocks onto the new weight. This occurs during both
increasing and decreasing weight values.
MANUAL (F35 = 2) – Press the F1 key before applying any weight to the scale. After the load has
stabilized, the display will hold the weight reading on the screen until the F1 key is pressed again.
The percentage hold setting (F36) unlocks the held reading and relocks on the new weight for
increasing weights only.
PEAK HOLD (F35 = 3) – The display updates as the load increases but not as the load decreases.
The value shown on the screen is the maximum weight applied to the scale. Press the F1 key to
activate peak hold mode. Press F1 again to exit peak hold mode and clear the peak value.
Use the Over and Under keys to program in a target weight for checkweighing.
CHECKWEIGHING APPLICATION (F30 = 6)
This mode is used to visually show a user if the materials applied to the scale meets, exceeds or
falls short of the target weight. These conditions correspond to ACCEPT, OVER and UNDER
respectively.
All vales are set thru use of the front panel. There are two ways to program the target weight values.
A password is required for each method:
OVER/UNDER – The upper and lower limits for the ACCEPT range are programmed into the scale
thru the corresponding keys on the front panel.
Once the values are programmed, each vertical bar on the LED represents one scale graduation,
e.g. if two LED bars are lit in the UNDER range, then you are two scale graduations away from
achieving ACCEPT.
TARGET – The target weight is derived from the weight applied to the scale platform. Place the
target weight onto the scale platform and then press the F1 key to store it.
Once the values are programmed, each vertical bar on the LED represents one scale graduation,
e.g. if two LED bars are lit in the UNDER range, then you are two scale graduations away from
achieving ACCEPT.