19
KEY-IN ZERO CALIBRATION VALUE (F19)
Note: This procedure is intended for emergency use only in the case of non-volatile memory loss.
A valid zero calibration value, obtained from a successful F16 calibration procedure, must
be used.
1. While in the Setup mode, scroll to "F 19" and then scroll down once using the ZERO key. The
display will momentarily show "CAL 0", followed by a value of zero
2. Use the front panel keys to key-in the actual value. If you make a mistake, use the C/CE key as a
backspace key.
3. After entering the exact value, press the MODE key to save the value. The display will show
"E CAL 0" momentarily, and then revert back up to F19.
KEY-IN SPAN CALIBRATION VALUE (F20)
Note: This procedure is intended for emergency use only in the case of non-volatile memory loss.
Valid span calibration values, obtained from a successful F17 calibration procedure, must be
used.
1. While in the Setup mode, scroll to "F 20", and then scroll down once using the ZERO key. The
indicator will prompt you to enter the information in Table 1 or the calibration report.
2. Use the front panel keys to key-in the actual value. If you make a mistake, use the C/CE key as a
backspace key.
3. After setting the exact value, press the MODE key to save the value. If the values are entered
successfully, the display will show "E" momentarily before continuing to the next parameter.
4. Repeat Steps 2 and 3 as needed until you reach the completion of the sequence, at which point
the indicator will revert back up to F20.