3.3.3 37°C Program (Fig 1.2)
This mode is the pre-programmed default 37°C incubation cycle. In this mode the incubator will
run an 18 hour incubation cycle at 37 °C. The timer can be altered to any setting between 0 and 24
hours. Adjusting the incubation time is described in section 3.3.7.
Pressing and holding the middle switch for 5 seconds will start the incubation cycle and the display
will change to the following:
The asterisk preceding the temperature will switch on and off indicating the status of the heater.
To stop or pause the incubation cycle press and hold the middle switch again for 5 seconds. The
display will return to the one shown in Fig 1.2.
Once the chamber has reached the pre-set temperature the timer will start to countdown. If for
any reason the temperature should drop, such as power failure, the timer will auto pause until the
temperature is back to the pre-set temperature. It should also be noted that if there is a power
failure the timer will reset back to the nearest hour. So a reset at 17:35 would result in a reset to
17:00.
The timer displayed is a countdown timer so it will display the time left not the elapsed time.
When the timer reaches 00:00 a buzzer will sound intermittently and the display will display done
A where A is the chamber.
20.9°C 24:00
->37° MEM
10.5v
*20.9°C 24:00
->37° RUN!
10.5v
Содержание AquaSafe WSL25 Plus
Страница 1: ...AquaSafe WSL25 Plus Water Safety Laboratory Instruction Manual...
Страница 7: ...Section 3 AquaSafe Micro biological Incubator Instruction Manual...
Страница 25: ...Section 4 AquaSafe Membrane Filtration Instruction Manual...
Страница 41: ...Section 5 AquaSafe Turbidity Tube Instruction Manual...
Страница 44: ...Section 6 AquaSafe Comparator Instruction Manual...
Страница 58: ...Section 7 HydroLite HL101 Pocket pH Tester Instruction Manual...
Страница 73: ...Section 8 HydroLite HL102 Pocket EC Tester Instruction Manual...