WARNING
:
The Reset function will restore the unit back to factory default settings; all calibration values and
selected parameters will be reset.
10. Press CAL key to enable or disable the reset function. Press ENTER key to confirm; the unit returns
to measurement mode.
EXIT THE SETUP MENU:
During the setup mode, to exit the setup menu, press ON/OFF key; the unit will return to measurement
mode immediately.
Enter