3.5 Rinse the pH electrode with distilled water again. Immerse the electrode into the pH10.01 (or
pH9.18) buffer solution. Stir the solution gently.
3.6 Press ENTER key to confirm. Wait for the measured value to stabilise; the display shows electrode
slope and “END”. Calibration is completed.
EXIT THE CALIBRATION:
During the calibration process, if you want to exit calibration, press ON/OFF key, the unit will return to
measurement mode immediately.
Enter