Electrode Care and Maintenance
•
After each use, the conductivity electrode should be rinsed thoroughly in deionised water.
•
If there is a build-up of solids inside the measurement area of the cell, these should be removed very
carefully with a cotton bud soaked in solvent, taking care not to touch the metal parts of the inner cell.
Electrode Replacement:
When the unit fails to calibrate or gives fluctuating readings for calibration standards, you need to
replace the electrode module.
1. Twist the electrode collar counter clockwise, pull the old electrode module away from the unit.
2. Align the slot on the new electrode module, gently push the module into the unit.
3. Twist the electrode collar clockwise until it is tight. Installation is completed.
Replacing the Batteries:
If the battery indicator disappears during the use, the batteries require replacing.
1. Twist the electrode collar counter clockwise, pull the electrode module out from the unit.
2. Insert two “AAA” batteries into the battery compartment (note polarity).
uS
Hold
Meas
+
Содержание AquaSafe WSL25 Plus
Страница 1: ...AquaSafe WSL25 Plus Water Safety Laboratory Instruction Manual...
Страница 7: ...Section 3 AquaSafe Micro biological Incubator Instruction Manual...
Страница 25: ...Section 4 AquaSafe Membrane Filtration Instruction Manual...
Страница 41: ...Section 5 AquaSafe Turbidity Tube Instruction Manual...
Страница 44: ...Section 6 AquaSafe Comparator Instruction Manual...
Страница 58: ...Section 7 HydroLite HL101 Pocket pH Tester Instruction Manual...
Страница 73: ...Section 8 HydroLite HL102 Pocket EC Tester Instruction Manual...