73
REFERENCE MATERIAL
ELECTROMAGNETIC COMPATIBILITY
This product conforms to the EMC standard (IEC 60601-1-2 : 2001).
a) MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC and needs
to be installed and put into service according to the EMC information provided in the ACCOM-
PANYING DOCUMENTS.
b) Portable and mobile RF communications equipment can affect MEDICAL ELECTRICAL
EQUIPMENT.
c) The use of ACCESSORIES, transducers and cables other than those specified, with the
exception of transducers and cables sold by the manufacturer of the EQUIPMENT or SYSTEM
as replacement parts for internal components, may result in increased EMISSIONS or
decreased IMMUNITY of the EQUIPMENT or SYSTEM.
d) The EQUIPMENT or SYSTEM should not be used adjacent to or stacked with other equip-
ment. If adjacent or stacked use is necessary, the EQUIPMENT or SYSTEM should be
observed to verify normal operation in the configuration in which it will be used.
e) The use of the ACCESSORY, transducer or cable with EQUIPMENT and SYSTEMS other than
those sepecified may result in increased EMISSION or decreased IMMUNITY of the EQUIP-
MENT or SYSTEM.
Item
Article code
Model No.
Length (m)
CBL_D1X-TRC-D15
445104200
—
5.0
CABLE_VS200-D2
445102300
—
1.0
CBL_D1X-TRC-VS2
445105200
—
5.0
Guidance and manufacturer's declaration - electromagnetic emissions
The TRC-50DX is intended for use in the electromagnetic environment specified below.
The customer or the user of the TRC-50DX should assure that it is used in such an environ-
ment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The TRC-50DX uses RF energy only for its internal
function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby elec-
tronic equipment.
RF emissions
CISPR 11
Class B
The TRC-50DX is suitable for use in all establish-
ments, including domestic establishments and those
directly connected to the public low-voltage power sup-
ply network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC61000-3-3
Complies
Содержание TRC-50DX
Страница 1: ...INSTRUCTION MANUAL RETINAL CAMERA TRC 50DX TRC 50DX ...