26
SETUP
FILTER IN/OUT
Decide the Ba (barrier) filter or Ex (exciter) filter link operation method.
"OUT" is the factory default.
1
Press the
switch on the "SET MENU" display.
The "FILTER IN/OUT" selection display will appear.
2
Press and select
or
. (Refer to the table below.)
3
Press the
switch, to store the data. The "SET
MENU" display will appear again.
If you press the
switch, the data will not be
stored. The data is reset as before and the "SET MENU"
display will appear again.
When you set the link operation of the Ba (barrier) filter and the
,
the relevant units will operate as shown in the following table because of the relation
between the setting for "FILTER IN/OUT" on the "SET MENU" display and the
switch on the control panel.
The
switch is displayed when the filter for fluorescein photography is set.
FILTER switch
ON
(The background of switch is black.)
OFF
(The background of switch is white.)
SET MENU
FILTER
IN/OUT
IN
Press the
,
and the Ba (barrier) filter is inserted
and it stays there.
The Ba (barrier) filter is not linked
with the
.
If you press the Ba (barrier) filter
switch, photography with blue filter is
possible.
OUT
Each time you press the
, the Ba (bar-
rier) filter is inserted and removed.
FILTER IN/OUT
IN
OUT
MEMORIZE
RETURN
PHOTOGRAPHY SWITCH
FILTER
PHOTOGRAPHY SWITCH
PHOTOGRAPHY SWITCH
PHOTOGRAPHY SWITCH
FILTER