BASIC OPERATIONS
6
Adjust the chinrest height by turning the chinrest adjusting knob so that the outside corner
of the patient's eye is level with the Canthus marker on the chinrest column. Then, let the
patient rest his/her forehead on the forehead rest. Use the headband if necessary.
(The headband is installed only in the relevant products.)
7
Guide the patient's eye so that the target part of the fundus can be photographed.
Guide the patient's eye appropriately by moving the external fixation target.
If it is difficult to fix the patient's eye on the target due to myopia or others, turn ON the
"FLICK" (External fixation blinking selection) on the "SET MENU" display to blink the exter-
nal fixation target for fixing the patient's eye more easily.
Refer to "FLICK (External fixation blinking selection)" on page 25.