SPECIFICATIONS AND PERFORMANCE
OPERATION PRINCIPLE
The observation light emitted from the illumination optical system illuminates the patient's eye
(the eye to be photographed), and the image formed by the observation/photography optical
systems is observed. By operating the photography switch on the instrument, the photography
light is emitted from the illumination optical system to illuminate the patient's eye. The image
formed by the observation/photography optical systems is photographed and recorded by the
35mm film camera (a product in the market) or an electronic photography device (a product in
the market).
The following filters, which can be installed/removed, are built in the illumination optical sys-
tem: One is the Ex filter and this selects and transmits the light that excites the fluorescent col-
oring matter in FAG photography. The other is the ICG-Ex filter and this selects and transmits
the infrared light that excites the fluorescent coloring matter in ICG fluorescein photography.
(The ICG-Ex filter is built in only Type IA.)
The following filters are built in the observation/photography optical systems: One is the Ba fil-
ter and this selects and filtrates the fluorescent wavelength in FAG photography. The other is
the ICG-Ba filter and this selects and filtrates the fluorescent wavelength in ICG fluorescein
photography. (The ICG-Ba filter is built in only Type IA.)
The angle of coverage can be changed by changing the lens located in the observation/pho-
tography optical systems. The illumination aperture located in the illumination optical system is
changed in the link operation with the change of the angle of coverage. So, the illumination
level is adjusted properly for photography.
The photography aperture located in the observation/photography optical systems is changed
properly for photography according to the photography technique.