64
OPTIONAL ACCESSORIES
Installing/removing the AUTO FLUO filter attachment
How to install the AUTO FLUO Ex filter
1
Turn the
OFF (
) and unplug the power cord.
2
Turn the control lever to raise the instrument body to its limit.
3
Remove the lamp house cover screw. Then, push the bottom of the lamp house cover and
pull it forward to remove it.
4
Set the
to "F1".
At this time, the filter frame "F1" is positioned at the lowest section of the turret.
The oil adheres on the outer circumference of the turret. Do not touch it.
5
To take out the filter frame easily, turn the
in approx. 45° counter-
clockwise.
6
Hold the projection of the filter frame with radio pliers,
etc. and pull it out straight.
7
Hold the projection of the frame on the AUTO FLUO
Ex filter (optional accessory) with radio pliers, etc. and
put it into the turret.
It is not necessary to distinguish between the front
and rear of the filter.
Set the filter frame to the innermost correctly.
Turn the
to make sure that it is not caught by anything.
Be careful not to damage or stain the filter with fingerprint.
8
Set the lamp house cover in the reversed procedure of the removal.
CAUTION
To avoid electric shock, be sure to turn the power switch off and
unplug the power cord before replacing the filter.
CAUTION
To avoid burns due to the hot filter unit, do not replace the filter with a
new one immediately after the power goes off.
NOTE
To prevent the glass from whitening due to fingerprints, do not touch
the filter with bare fingers.
NOTE
Since the filter is not resistible to shocks, handle it with particular care.
POWER SWITCH
FILTER SWITCHING KNOB
FILTER SWITCHING KNOB
FILTER SWITCHING KNOB